Chapter 3 Constant Current Tests RIGOL
DP800 Performance Verification Manual
3-9
Test Record Forms:
Table 3-6 DP831A CC Linear Regulation Rate Test Record Form
Channel CH1 (8V/5A) CH2 (30V/2A) CH3 (-30V/2A)
Specification <750μA <450μA <450μA
DP831A
Setting
Voltage 8.4V 32V -32V
Current 5A 2A 2A
Measurement
Result
U
0
U
1
U
2
|U
1
-U
0
|/R
M
[1]
|U
2
-U
0
|/R
M
[1]
Note
[1]
: R
M
=0.022Ω.
Table 3-7 DP832A CC Linear Regulation Rate Test Record Form
Channel CH1 (30V/3A) CH2 (30V/3A) CH3 (5V/3A)
Specification <550μA
DP832A
Setting
Voltage 32V 32V 5.3V
Current 3A 3A 3A
Measurement
Result
U
0
U
1
U
2
|U
1
-U
0
|/R
M
[1]
|U
2
-U
0
|/R
M
[1]
Note
[1]
: R
M
=0.022Ω.
Table 3-8 DP832 CC Linear Regulation Rate Test Record Form
Channel CH1 (30V/3A) CH2 (30V/3A) CH3 (5V/3A)
Specification <550μA
DP832
Setting
Voltage 32V 32V 5.3V
Current 3A 3A 3A
Measurement
Result
U
0
U
1
U
2
|U
1
-U
0
|/R
M
[1]
|U
2
-U
0
|/R
M
[1]
Note
[1]
: R
M
=0.022Ω.

Other manuals for Rigol DP800A Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DP800A Series and is the answer not in the manual?

Rigol DP800A Series Specifications

General IconGeneral
BrandRigol
ModelDP800A Series
CategoryPower Supply
LanguageEnglish

Related product manuals