RIGOL  Chapter 2 Calibration Process 
2-2    DP800 Calibration Guide 
3.  Meas Val 
Press this softkey after selecting a calibration point. The Meas Value column of 
the corresponding point becomes editable (the reference voltage or current of 
the corresponding point is displayed on the interface) and Input Val becomes 
available. At this point, use the numeric keyboard or the knob and direction keys 
to input the measurement value. 
 
4.  Input Val 
After selecting a calibration point and inputting the measurement value, press 
this softkey to write the measurement value into the system and finish the 
calibration of the point selected. 
 
5.  Save 
Press this softkey to save the current calibration values and exit the calibration 
interface. 
 
6.  Sign 
This softkey is used to control the sign of the data input using the numeric 
keyboard. The data input is a positive value when the sign is set to “+” and will 
be converted to a negative value automatically when the sign is set to “-”. 
 
 
The calibration method of each calibration item is introduced in detail in the following 
part of this chapter.  
 
 
 
CAUTION 
You are recommended to calibrate each item following the sequence 
below.