RIGOL Chapter 2 Calibration Process
2-4 DP800 Calibration Guide
Note: The differences between the V DAC calibration steps of the channels which
support the Sense function (CH2 of DP821A and DP811A) and the above calibration
steps are as shown below.
1) The following operations should be performed before the first step (connect the
channel output terminals of DP800 with the voltage input terminals of the digital
multimeter).
Short-circuit the positive terminal (+) of the channel output terminals with
the positive terminal (S+) of the Sense terminals of the corresponding
channel using the short circuitor.
Short-circuit the negative terminal (-) of the channel output terminals with
the negative terminal (S-) of the Sense terminals of the corresponding
channel using the short circuitor.
2) Turn on the Sense function of the corresponding channel before the second step
(enable the channel output).