  Pass/Fail: 
The instrument will output a negative pulse via this connector when a failed 
waveform is detected during the pass/fail test. The instrument continuously 
outputs a low level via this connector when no failed waveform is detected. 
 
5. Source Output 
The output terminals of the built-in dual-channel source of the oscilloscope. 
When Source1 or Source2 is enabled, the signal currently set can be output 
through the [Source1] or [Source2] connector on the rear panel.  
 
6. Lock Hole 
You can lock the instrument to a fixed location by using the security lock (please 
purchase it yourself) via the lock hole.  
 
7. Fuse 
If a new fuse is required, please use the specified fuse (250V, T2A). The 
operation method is as follows: 
a)  Turn off the instrument, disconnect the power, and then remove the power 
cord. 
b)  Insert a small straight screwdriver into the slot at the power cord connector 
and pry out the fuse holder gently.  
c)  Take out the fuse and replace it with a specified fuse. Then, install the fuse 
holder. 
 
8. AC Power Cord Connector 
AC power input terminal. The power requirements of this oscilloscope are 
100-240 V, 45-440 Hz. Use the power cord provided with the accessories to 
connect the instrument to AC power. Then, you can press the Power key on the 
front panel to start the instrument.