RIGOL Chapter 2 Performance Verification Test
2-22 RSA5000 Performance Verification Guide
Input Attenuation Switching Uncertainty Test
Specification
Input Attenuation Switching Uncertainty
fc = 50 MHz, relative to 10 dB, PA off, 20 to 30
Switching Uncertainty
<0.3 dB
Test Devices
1. Signal Generator × 1
2. Power Meter × 1
3. Power Sensor × 1
4. Power Meter Connecting Cable × 1
5. Program-controlled Attenuator:
Attenuator/Switch Driver × 1
Step Attenuator (Step: 1 dB) × 1
Step Attenuator (Step: 10 dB) × 1
Attenuator Interconnect Kit × 1
6. Dual-N Cable × 2