RIGOL Appendix
2 RSA5000 Performance Verification Guide
SSB Phase Noise Test:
Output Frequency
of Signal Generator
Offset
Measurement
Result
Limit
Pass/Fail
500 MHz
10 kHz offset
<-106 dBc/Hz, <-108 dBc/Hz
(typical)
Absolute Amplitude Accuracy Test:
Power Meter
Measurement
Value A1
Spectrum
Analyzer
Measurement
Value A2
Calculation
Result (A1 - A2)
Limit
Pass/Fail
<0.3 dB
Frequency Response Test (PA Off):
Reference
Value 1
Reference
Value 2
Signal
Generator
Output
Frequency
Power Meter
Measurement
Value A1
Spectrum
Analyzer
Measurement
Value A2
System
Error
[1]
Global
Error
[2]
Frequency
Response
[3]
Limit
Pass/Fail
100 kHz
100
kHz to
3.2
GHz:
<0.5 dB,
<0.3 dB
(typical)
3.2 GHz
to 6.5
GHz:
<0.7 dB,
<0.5 dB
(typical)
500 MHz
2 GHz
3 GHz
4 GHz
5 GHz
6 GHz
6.499995
GHz