3 Product Components 3-1. Main unit and standard accessories
GX-8000 - 8 -
3
Product Components
3-1. Main unit and standard accessories
After opening the package, check the main unit and accessories.
If anything in the following list is not included, contact RIKEN KEIKI.
<Main Unit>
<Standard Accessories>
AA alkaline battery: 3
Gas sampling probe
and gas sampling hose: 1
Shoulder strap: 1
Operating manual
Product warranty
GX-8000
main unit
Dry battery unit (BUD-8000 (G))
DANGER
About explosion-proof
Do not modify or change the circuit or structure, etc.
When using the gas monitor in a hazardous area, take the following countermeasures for
preventing dangers resulting from electrostatic charges.
(1) Wear anti-static clothes and conductive shoes (anti-static work shoes).
(2) For indoor use, use the gas monitor while standing on a conductive work floor (with a
leakage resistance of 10 M or less).
The battery units that can be connected are the BUL-8000(G) (certificate number TC19437) or
BUD-8000(G) (certificate number TC19438).
The specifications of the gas monitor are as follows:
Pump circuit: Allowable voltage of 4.95 V, allowable current of 1.12 A, and
allowable power of 1138 mW
Combustible gas sensor circuit: Allowable voltage of 4.95 V, allowable current of 0.834 A, and
allowable power of 853 mW
Buzzer circuit: Allowable voltage of 4.95 V, allowable current of 0.431 A, and
allowable power of 441 mW
Main circuit: Allowable voltage of 4.95 V, allowable current of 0.717 A, and
allowable power of 733 mW
Backup circuit: 3.0 VDC, 10 μA

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Riken Keiki GX-8000 and is the answer not in the manual?

Riken Keiki GX-8000 Specifications

General IconGeneral
DisplayLCD with backlight
Operating Humidity0 to 95% RH (non-condensing)
Ingress ProtectionIP67
CommunicationUSB
Gas TypeO2, H2S, CO
Detection PrincipleCatalytic combustion, Electrochemical cell
Power SourceRechargeable battery
Operating Temperature-20°C to 50°C (-4°F to 122°F)
Alarm Set PointsUser adjustable
Response TimeWithin 30 seconds
CertificationsATEX, IECEx, CSA
AlarmsVisual, Audible, and Vibration
Data LoggingYes
Detection RangeCombustibles: 0-100% LEL, CO: 0-500 ppm, H2S: 0-100 ppm