Command Answer Description
S01; Select unit 1
IAD?1; 1,3000,0,1,0,500,1 CRLF Current settings are:
max1 = 3000,
no decimal point,
e1 = 1,
x10 mode is off,
additive tare is 500,
sense line check is on
IAD1,4000,1,2,0,200,0; 0 CRLF New settings are:
max1 = 4000,
1 digit after decimal point,
e1 = 2,
x10 mode is off,
additive tare is 200,
sense line check is off
TDD1; 0 CRLF Save setting
If IAD? is issued without the range parameter then the returned data is range 1 for single range setup or range
2 for dual interval or dual range setup. In this way it is possible to query the maximum load without the need
to issue a WMD? Command to determine the weighing mode.
21.3.23 ICR: Measurement Rate
Set the fundamental measurement frequency of the instrument.
General
No. of parameters 1
Save changes With TDD1
Increment Trade Counter Yes
Parameter Details
Parameter Description Range Default
1 Measurement Rate in Hertz 10 = 10Hz
12 = 12.5Hz
15 = 15Hz
20 = 20Hz
25 = 25Hz
30 = 30Hz
50 = 50Hz
60 = 60Hz
100 = 100Hz
50
Note
Frequencies other than those listed above will be accepted but the indicator will
operate at the nearest frequency in the table.
Example
Command Answer Description
S01; Select unit 1
ICR?; 50 CRLF Query current measurement rate
ICR60; 0 CRLF Change to 60 Hz
TDD1; 0 CRLF Save setting
136 C500-600-2.0.3

Table of Contents

Other manuals for Rinstrum C520

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rinstrum C520 and is the answer not in the manual?

Rinstrum C520 Specifications

General IconGeneral
BrandRinstrum
ModelC520
CategoryTouch Panel
LanguageEnglish

Related product manuals