Document number R2019-10-01 Span Calibration Guide
Revision 02
Page 11 of 16
www.ritchie-uk.com
3.3 ID 5000 - Calibration
3.3.1 Screw the load bars connectors securely to the correct positions
3.3.2 Turn on indicator with
power button.
3.3.3 Start at the home
screen, select the
settings page, by
using the grey button
at the side.
3.3.4 Select Weighing settings, by using the grey button at the
side
Load bar connectors