Document number R2019-10-01 Span Calibration Guide
Revision 02
Page 13 of 16
www.ritchie-uk.com
3.3.9 Enter the test weight into ID5000, add test weight
to weigh scale, keeping central for best accuracy.
3.3.10 Press enter to start span calibration. DO NOT touch
weigh scale while calibration is being carried out.
Screen will display “Test weight measurement is
done”
3.3.11 Remove weight when completed.
3.3.12 Return to weighing screen, using the back key
.
Test weigh scale with known weight to verify