Command Reference
R&S
®
CMW500
203User Manual 1173.9463.02 ─ 02
Example:
CONFigure:BASE:FDCorrection:CTABle:CREate
'mytable', 1900000000, 0.5, 2000000000, 0.7
Create the table 'mytable' with two entries: 0.5 dB at 1900 MHz
and 0.7 dB at 2000 MHz
Usage: Setting only
Firmware/Software: V1.0.5.3
CONFigure:BASE:FDCorrection:CTABle:ADD '<TableName>', <Frequency_0>,
<Attenuation_0>[, ..., <Frequency_n>, <Attenuation_n>]
Adds entries to an existing correction table. At least one parameter pair has to be speci-
fied. A command with an incomplete pair (e.g. <Frequency> without <Attenuation>) is
ignored completely.
You can add parameter pairs in any order. The table entries (pairs) are automatically
sorted from lowest to highest frequency.
Setting parameters:
'<TableName>' String parameter identifying the table. To display a list of existing
tables use the command
CONFigure:BASE:FDCorrection:CTABle:CATalog?.
<Frequency...> Range: 100E+6 Hz to 6E+9 Hz
Increment: 0.1 Hz
Default unit: Hz
<Attenuation...> Range: -50 dB to 90 dB
Increment: 0.01 dB
Default unit: dB
Example:
CONFigure:BASE:FDCorrection:CTABle:ADD
'mytable', 1925000000, 0.55, 1975000000, 0.65
Add two entries to the table 'mytable': 0.55 dB at 1925 MHz
and 0.65 dB at 1975 MHz
Usage: Setting only
Firmware/Software: V1.0.5.3
CONFigure:BASE:FDCorrection:CTABle:DELete '<TableName>'
Deletes a correction table from the hard disk.
Setting parameters:
'<TableName>' String parameter identifying the table. To display a list of existing
tables use the command
CONFigure:BASE:FDCorrection:CTABle:CATalog?.
Example:
CONFigure:BASE:FDCorrection:CTABle:DELete
'mytable'
Usage: Setting only
Instrument-Control Commands