Command Reference
R&S
®
CMW500
204User Manual 1173.9463.02 ─ 02
Firmware/Software: V1.0.5.3
CONFigure:BASE:FDCorrection:CTABle:CATalog?
Returns a comma separated list of table names. All correction tables currently stored on
the hard disk are listed.
Example:
CONFigure:BASE:FDCorrection:CTABle:CATalog?
Returns: 'mytable','setup 1','setup 3G'.
Usage: Query only
Firmware/Software: V1.0.5.3
CONFigure:BASE:FDCorrection:CTABle:DETails? '<TableName>'[, <Start>[,
<Count>]]
Returns the entries of a correction table. Each entry consists of a frequency [Hz] and an
attenuation [dB].
Query parameters:
'<TableName>' String parameter identifying the table. To display a list of existing
tables use the command
CONFigure:BASE:FDCorrection:CTABle:CATalog?.
<Start> Index number of the first entry to be listed. The first entry of a table
has index number 0.
Default: 0
<Count> Maximum number of entries to be listed. By default all entries from
<Start> to the end of the table are listed.
Example:
CONFigure:BASE:FDCorrection:CTABle:DETails?
'mytable', 3, 1
Returns the entry with index number 3, e.g.: 2000000000,0.7
Usage: Query only
Firmware/Software: V1.0.5.3
CONFigure:BASE:FDCorrection:CTABle:LENGth? '<TableName>'
Returns the number of entries (i.e. pairs of frequency and attenuation) of a correction
table.
Query parameters:
'<TableName>' String parameter identifying the table. To display a list of existing
tables use the command
CONFigure:BASE:FDCorrection:CTABle:CATalog?.
Example:
CONFigure:BASE:FDCorrection:CTABle:LENGth?
'mytable'
For 'mytable' containing 4 frequency/attenuation pairs, the
result is 4.