Command Reference
R&S
®
CMW500
205User Manual 1173.9463.02 ─ 02
Usage: Query only
Firmware/Software: V1.0.5.3
CONFigure:FDCorrection:ACTivate <Connector>, '<TableName>'[, <Direction>]
Activates a correction table for the signal path using a specific RF connector.
For bidirectional connectors the table can be applied to both directions or to one direction.
It is possible to assign different tables to the directions of a bidirectional connector, see
example.
Depending on the installed hardware and the active sub-instrument only a subset of the
listed connector values is allowed. The mapping of virtual connector names to physical
connectors also depends on the active sub-instrument. For details see Signal Path Set-
tings.
Setting parameters:
<Connector> RF1C | RF1O | RF2C | RF3C | RF3O | RF4C | RFAO | RFAC |
RFBC
RF1C, RF2C, RF3C, RF4C, RF1O, RF3O:
RF 1 COM to RF 4 COM and RF 1/3 OUT front panel connectors
RFAC, RFBC, RFAO:
Virtual names for the RF COM and RF OUT connectors
'<TableName>' String parameter identifying the table. To display a list of existing
tables use the command
CONFigure:BASE:FDCorrection:CTABle:CATalog?.
<Direction> RXTX | RX | TX
Specifies the direction to which the correction table shall be
applied. RX means input and TX means output. For a pure output
connector RX is ignored.
RXTX: both directions (for output connector only output)
RX: input (no effect for output connector)
TX: output
Default: RXTX
Example:
CONFigure:FDCorrection:ACTivate RF1C,
'mytable_in', RX
CONFigure:FDCorrection:ACTivate RF1C,
'mytable_out', TX
Different tables are activated for the input and output direction of
the connector RF1C.
Usage: Setting only
Firmware/Software: V1.0.5.3
CONFigure:FDCorrection:DEACtivate <Connector>
Deactivates any correction tables for a specific RF connector.
Instrument-Control Commands