Remote Commands in the Receiver Application
R&S
®
ESW
505User Manual 1177.6298.02 ─ 08
Suffix:
<n>
.
irrelevant
<m> irrelevant
Parameters:
<TraceNumber> 1 to 6
Range: 1 to 6
*RST: 1
Manual operation: See "Signal Tracking" on page 178
See "IF Signal Tracking" on page 181
DISPlay[:WINDow<n>]:TRACe<t>:X:SPACing
<Scale>
This command selects the scale of the x-axis.
Suffix:
<n>
.
irrelevant
<t> irrelevant
Parameters:
<Scale> LINear
Linear scale of the frequency axis.
LOGarithmic
Logarithmic scale of the frequency axis.
*RST: LOGarithmic
Example: //Select a linear scale for the x-axis
DISP:TRAC:X:SPAC LIN
Manual operation: See "Frequency Axis Scale" on page 179
[SENSe:]FREQuency:CENTer <Frequency>
This command defines the measurement frequency for measurements in the frequency
or time domain.
Parameters:
<Frequency> <numeric value>
Numeric value in Hz.
Range: Refer to the datasheet
*RST: fmax / 2
Default unit: Hz
Example: //Define receiver frequency
FREQ:CENT 100MHz
Manual operation: See "Receiver Frequency" on page 178