Remote Control
R&S
®
FPC
203User Manual 1178.4130.02 ─ 07
*OPC............................................................................................................................204
*OPT?...........................................................................................................................204
*RST.............................................................................................................................204
*SRE............................................................................................................................ 204
*STB?...........................................................................................................................205
*TRG............................................................................................................................ 205
*TST?...........................................................................................................................205
*WAI.............................................................................................................................205
*CLS
Clear status
Sets the status byte (STB), the standard event register (ESR) and the EVENt part of
the QUEStionable and the OPERation registers to zero. The command does not
alter the mask and transition parts of the registers. It clears the output buffer.
Usage:
Setting only
*ESE <Value>
Event status enable
Sets the event status enable register to the specified value. The query returns the con-
tents of the event status enable register in decimal form.
Parameters:
<Value> Range: 0 to 255
*ESR?
Event status read
Returns the contents of the event status register in decimal form and then sets the reg-
ister to zero.
Return values:
<Contents> Range: 0 to 255
Usage: Query only
*IDN?
Identification
Returns the instrument identification.
Return values:
<ID> "Rohde&Schwarz,<device type>,<part number>/<serial num-
ber>,<firmware version>"
Example:
Rohde&Schwarz,FPC1000,1328.6660K02/xxxxxx,1.00
Usage: Query only
Common Commands

Table of Contents

Other manuals for Rohde & Schwarz FPC Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rohde & Schwarz FPC Series and is the answer not in the manual?

Rohde & Schwarz FPC Series Specifications

General IconGeneral
BrandRohde & Schwarz
ModelFPC Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals