Controlling the R&S FSVA/FSV Remotely
R&S
®
FSVA/FSV
168Quick Start Guide 1321.3066.02 ─ 06
Risk of network errors
Connection errors can affect the entire network. If your network does not
support DHCP, or if you choose to disable dynamic TCP/IP configuration,
you must assign valid address information before connecting the instrument
to the LAN. Contact your network administrator to obtain a valid IP address.
Assigning the IP address on the instrument
1. Press the SETUP key.
2. Press the "General Setup" softkey.
3. Press the "Network Address" softkey.
The submenu is displayed.
4. Toggle the "DHCP On/Off" softkey to the required mode.
If DHCP is "Off", you must enter the IP address manually, as described in the
following steps.
Note: When DHCP is changed from "On" to "Off", the previously set IP
address and subnet mask are retrieved.
If DHCP is "On", the IP address of the DHCP server is obtained automatically.
The configuration is saved, and you are prompted to restart the instrument.
You can skip the remaining steps.
Note: When a DHCP server is used, a new IP address may be assigned each
time the instrument is restarted. This address must first be determined on the
instrument itself. Thus, when using a DHCP server, it is recommended that
you use the permanent computer name, which determines the address via the
DNS server (see Chapter 8.6.3, "Using Computer Names", on page 170).
5. Press the "IP Address" softkey and enter the IP address, for example
192.0.2.0. The IP address consists of four number blocks separated by dots.
Every block contains a maximum of 3 numbers.
6. Press the "Subnet Mask" softkey and enter the subnet mask, for example
255.255.255.0. The subnet mask consists of four number blocks separated by
dots. Every block contains a maximum of 3 numbers.
If you have entered an invalid IP address or subnet mask, the message "out of
range" is displayed in the status line. The "Edit" dialog box remains open, and
you can start again. If the settings are correct, the configuration is saved, and
you are prompted to restart the instrument.
Setting Up a Network (LAN) Connection