EasyManua.ls Logo

Rohde & Schwarz FSW Series - 5.4 Input and Frontend Settings

Rohde & Schwarz FSW Series
123 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Tour
R&S
®
FSW
55Getting Started 1338.4102.02 ─ 08
The upper BNC connectors BASEBAND INPUT I and BASEBAND INPUT Q are
used to input:
Single-ended signals
The positive signal input for differential signals
Input from active Rohde & Schwarz probes (see data sheet)
The lower BNC connectors Ī and Ǭ are used to input the negative signal for differ-
ential signals.
R&S FSW85
The R&S FSW85 provides only two connectors; differential input is not sup-
ported.
Complex signal input (I+jQ)
For complex signal input (I+jQ), always use two identical cables for the I and
Q connectors (same length, same type, same manufacturer). Otherwise,
time delay or gain imbalance can occur between the different cables, which
cannot be calibrated.
All connectors have a fixed impedance of 50 Ω. Do not overload the input. For
maximum allowed values, see the data sheet.
Input via the Analog Baseband Interface can be enabled in the I/Q Analyzer, the
Analog Demodulation application, or in one of the optional applications that proc-
ess I/Q data (where available).
For more information on the Analog Baseband Interface, see the R&S FSW I/Q
Analyzer and I/Q Input User Manual.
Rear Panel View

Table of Contents

Other manuals for Rohde & Schwarz FSW Series

Related product manuals