Operating the Instrument
R&S
®
FSW
91Getting Started 1338.4102.02 ─ 08
7.1.3 Marker Information
Marker information is provided either in the diagram grid or in a separate marker
table, depending on the configuration.
Marker information in diagram grid
Within the diagram grid, the x-axis and y-axis positions of the last two markers or
delta markers that were set are displayed, if available, as well as their index. The
value in the square brackets after the index indicates the trace to which the
marker is assigned. (Example: M2[1] defines marker 2 on trace 1.) For more than
two markers, a separate marker table is displayed beneath the diagram by
default.
Marker information in marker table
In addition to the marker information displayed within the diagram grid, a separate
marker table may be displayed beneath the diagram. This table provides the fol-
lowing information for all active markers:
Type Marker type: N (normal), D (delta), T (temporary, internal), PWR
(power sensor)
Ref Reference (for delta markers)
Trc Trace to which the marker is assigned
X-value X-value of the marker
Y-value Y-value of the marker
Func Activated marker or measurement function
Func .Result Result of the active marker or measurement function
The functions are indicated with the following abbreviations:
FXD Fixed reference marker
PHNoise Phase noise measurement
CNT Signal count
TRK Signal tracking
NOIse Noise measurement