EasyManua.ls Logo

Rohde & Schwarz FSW26 - Page 52

Rohde & Schwarz FSW26
123 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Tour
R&S
®
FSW
52Getting Started 1338.4102.02 ─ 08
512 MHz (with option B512/U512)
1200 MHz (with option B1200/U1200)
2001 MHz (with option B2001/U2001)
4001 MHz (with option B4001/U4001)
6001 MHz (with option B6001/U6001)
8001 MHz (with option B8001/U8001)
While the extension can be activated or deactivated manually in the R&S FSW
base unit (I/Q Analyzer application), it is activated automatically in some applica-
tions that also support I/Q data analysis. See the application-specific documenta-
tion for details.
Together with the bandwidth extension an additional IF output connector is provi-
ded ("IF WIDE OUTPUT"). As opposed to the default IF/VIDEO/DEMOD OUT-
PUT connector, the IF output frequency of the optional connector cannot be
defined manually, but is determined automatically depending on the center fre-
quency. For details on the used frequencies, see the data sheet. The IF WIDE
OUTPUT connector is used automatically when the bandwidth extension is acti-
vated (i.e. for bandwidths > 80 MHz).
5.2.7 Digital I/Q 40G Streaming Output Connector (R&S FSW-
B517)
The Digital I/Q 40G Streaming Output (QSFP+) connector is provided by the
hardware of any bandwidth extension option for 512 MHz or more.
If necessary, remove the metal cover from the connector on the rear panel of the
R&S FSW.
Rear Panel View

Table of Contents

Other manuals for Rohde & Schwarz FSW26

Related product manuals