Measurements
R&S
®
FSW
218User Manual 1173.9411.02 ─ 19
Alpha ← Channel Power Settings
Sets the alpha value of the RRC filter (if activated).
Remote command:
[SENSe:]ESPectrum<sb>:FILTer[:RRC]:ALPHa on page 748
5.6.5.4 Power Classes
In the "Power Classes" tab of the "Spectrum Emission Mask" dialog box you configure
power classes for which you can then define sweep list ranges. For details see "Power
classes" on page 200.
Used Power Classes...................................................................................................218
PMin/PMax..................................................................................................................219
Sweep List...................................................................................................................219
Adding or Removing a Power Class........................................................................... 219
Used Power Classes
Defines which power classes are considered for the SEM measurement. Limits can be
defined only for used power classes. It is only possible to select either one specific
power class or all of the defined power classes.
If "All" is selected, the power class that corresponds to the currently measured power
in the reference range is used for monitoring. The limits assigned to that power class
are applied (see "Abs Limit Start/Stop <n>" on page 213 and "Rel Limit Start/Stop <n>"
on page 213).
Remote command:
CALCulate:LIMit:ESPectrum<sb>:PCLass<class>[:EXCLusive]
on page 752
To define all limits in one step:
CALCulate:LIMit:ESPectrum<sb>:PCLass<class>:LIMit[:STATe]
on page 752
Spectrum Emission Mask (SEM) Measurement