Measurements
R&S
®
FSW
245User Manual 1173.9411.02 ─ 19
For details on the reference level see chapter 6.4.1.1, "Reference Level",
on page 393.
Remote command:
[SENSe:]LIST:RANGe<range>:RLEVel on page 772
RF Att. Mode
Activates or deactivates the auto mode for RF attenuation.
For details on attenuation see chapter 6.4.1.2, "RF Attenuation", on page 394.
Remote command:
[SENSe:]LIST:RANGe<range>:INPut:ATTenuation:AUTO on page 770
RF Attenuator
Sets the attenuation value for that range.
For details on attenuation see chapter 6.4.1.2, "RF Attenuation", on page 394.
Remote command:
[SENSe:]LIST:RANGe<range>:INPut:ATTenuation on page 770
Preamp
Switches the preamplifier on or off.
For details on the preamplifier see "Preamplifier (option B24)" on page 399.
Remote command:
[SENSe:]LIST:RANGe<range>:INPut:GAIN:STATe on page 770
Sweep Points
Sets the number of sweep points for the specified range.
For details on sweep points see chapter 6.5.1.8, "How Much Data is Measured: Sweep
Points and Sweep Count", on page 407.
Remote command:
[SENSe:]LIST:RANGe<range>:POINts[:VALue] on page 772
Stop After Sweep
This command configures the sweep behavior.
"ON"
The R&S FSW stops after one range is swept and continues only if
you confirm (a message box is displayed).
"OFF"
The R&S FSW sweeps all ranges in one go.
Remote command:
[SENSe:]LIST:RANGe:BREak on page 767
Transducer
Sets a transducer for the specified range. You can only choose a transducer that fulfills
the following conditions:
The transducer overlaps or equals the span of the range.
The x-axis is linear.
The unit is dB.
Spurious Emissions Measurement