[SENSe:]PMETer<p>:UPDate[:STATe] <State>
This command turns continuous update of power sensor measurements on and off.
If on, the results are update even if a single sweep is complete.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<State> ON | OFF
*RST: OFF
Example:
PMET1:UPD ON
The data from power sensor 1 is updated continuously.
Manual operation: See "Continuous Value Update" on page 325
UNIT<n>:PMETer<p>:POWer <Unit>
This command selects the unit for absolute power sensor measurements.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Unit> DBM | WATT | W
*RST: DBM
Example:
UNIT:PMET:POW DBM
Manual operation: See "Unit/Scale" on page 326
UNIT<n>:PMETer<p>:POWer:RATio <Unit>
This command selects the unit for relative power sensor measurements.
Suffix:
<p>
.
1...4
Power sensor index
Parameters:
<Unit> DB | PCT
*RST: DB
Example:
UNIT:PMET:POW:RAT DB
Manual operation: See "Unit/Scale" on page 326