ACTIVE PROBE HZO30
All data valid at 23 °C after 30 minute warm-up
Electrical specications
Attenuation ratio: 10:1 ± 0.5 % at DC
Dynamic Measuring
Range: ±8 V
System bandwidth
1)
: 600 MHz (-3 dB)
Bandwidth (probe only): 1 GHz
Maximum rated
input voltage: 20 V
Max. Input Voltage and
Dynamic Measuring Range
The HZO30 Aktiv-Tastkopf active probe is pro-
tected against electro-static-discharge voltage
(ESD). Applying input-voltages outside the speci-
ed limits can result in destruction of the probe´s
amplier.
Make sure you comply with the below input
voltage vs. frequency chart to avoid input
linearity errors and damage to the probe.
1
10
100
1M 10M 100M 1G 10G
Voltage [V] AC pk
Frequency [Hz]
Input Voltage vs. Frequency
Dynamic Measuring Range
Maximum Input Voltage
Electrical characteristics
Input resistance (System):
>
1 MΩ
Input capacitance (System): 0.9 pF (typical)
Oscilloscope Input Coupling: 50Ω AC / DC
Overshoot: 6% (typical)
1) Connected to oscilloscope > 500 MHz
Prob e HZ O 30
Input Impedance
Note that the input impedance of the probe
decreases as the frequency of the applied
signal increases.
10k
100k
1M
10M
100 1k 10k 100k 1M 10M 100M 1G
Frequency [Hz]
Typical input impedance
I Z I [Ohm]
Mechanical characteristics
Weight (probe only): 96 g
Cable length: 1.3 m
Environmental specications
Altitude
operating: up to 2000 m
non-operating: up to 15000 m
Temperature range
operating: 0° C to +50° C
non-operating: -40° C to +70° C
Maximum relative humidity
operating: 80 %
for temperatures up to +30° C
decreasing linearly to 40 % at 50 °C