Known-Issues
External Reference
In rare cases the error message 'RFB6.BB: Output unleveled: Gain
Control on Upper Limit' appears if the external reference frequency
was switched from external to internal.
421671
1.3 Version 4.60.112.22
Released: May 2019
New Options
None
New Functionality / Changed Behavior
RF
Sweep
Sweep will be switched off in case of considerable overrun conditions.
456033
List Mode
List Mode will be switched off in case of considerable overrun
conditions.
456033
LF Output
A new value "Load Impedance" in certain emulations
542234
Fixed Issues
Common
General
Slide switches can be operated in remote state.
501226
Internal Adjustment
Improved Table Adjustment. Internal Adjustments have to be done.
562367
Remote Control
Error handling: SYSTEM:SERROR? delivers info-messages, different
from SYSTEM:ERROR?
555983
Save Recall
Frequency is recalled even in case of "Exclude Frequency"
453572
RF
AM modulation
The RF level is higher than expected.
510294
RF Level Sweep
In rare cases the level steps are different.
542768
Pulse Generator
The trigger mode 'Ext Single' doesn't work.
522691
1.4 Version 4.50.074.45
Released: Dezember 2018