Remote control commands
R&S
®
RTM3000
442User Manual 1335.9090.02 ─ 09
Parameters:
<State> ON | OFF
*RST:  OFF
17.4.4  Passive probes
PROBe<m>:SETup:ATTenuation:UNIT.............................................................................442
PROBe<m>:SETup:ATTenuation:MANual.........................................................................442
PROBe<m>:SETup:GAIN:UNIT.......................................................................................442
PROBe<m>:SETup:GAIN:MANual...................................................................................443
PROBe<m>:SETup:ATTenuation:UNIT <Unit>
Selects the unit that the probe can measure.
Suffix: 
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Unit> V | A
PROBe<m>:SETup:ATTenuation:MANual <ManualAttenuation>
Sets the attenuation of the probe.
Suffix: 
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<ManualAttenuation> Range:  0.0001  to  10e6
PROBe<m>:SETup:GAIN:UNIT <Unit>
Selects the unit that the probe can measure.
Suffix: 
<m>
.
1..4
Selects the input channel. The number of channels depends on
the instrument.
Parameters:
<Unit> V | A
Waveform setup