WGENerator:VOLTage:OFFSet <Offset>
Sets the DC offset of the selected generator function.
Parameters:
<Offset> Numeric value
Range:  -3.00000E+00  to  3.00000E+00
*RST:  0.00E+00
Default unit: V
WGENerator:FREQuency <Frequency>
Defines the frequency.
Parameters:
<Frequency> Range:  Depends on the selected function. For details, refer
to the datasheet.
*RST:  10 kHz
WGENerator:FUNCtion:PULSe:DCYCle <DutyCycle>
Defines the duty cycle value of the generator function pulse.
Parameters:
<DutyCycle> Numeric value
Range:  1.000E+01  to  9.000E+01
*RST:  2.500E+01
Default unit: %
WGENerator:TRIangle:SYMMetry <Symmetry>
Sets the symmetry for the triangle waveform.
Parameters:
<Symmetry> 50% defines symmetric triangles. Values <50% define triangles
with steeper rising edge leaned to the left. Values >50% define
triangles with steeper falling edge leaned to the right.
Range:  1  to  99
Increment:  1
*RST:  50
Default unit: %
WGENerator:FUNCtion:PULSe:ETIMe <EdgeTime>
Sets the pulse edge time.
Parameters:
<EdgeTime>
Signal generation (option R&S RTM-B6)