HYGRO PALM HP32
Short Instruction Manual
1 GENERAL DESCRIPTION
The HP32 is a highly accurate handheld instrument compatible with all HC2 and HC2A probes
to perform measurement and logging of relative humidity, temperature and calculated psychro-
metric parameters.
1.1 PACKAGE CONTENT
HygroPalm HP32
Soft Case
Micro-USB Service Cable (AC0006)
Belt Buckle Set
Lanyard
Certicate
HW4-LITE license key
2 ASSEMBLY AND FIXATION
2.1 PROBE
The HP32 is compatible with all HC2 and HC2A probes. Simply connect it to the probe input
on top.
A calibration of the HP32 after changing the probe is not necessary.
2.2 TRIPOD SOCKET
The standardized tripod socket (1/4”-20 UNC thread) at the back of the device enables a wide
range of conventional mounting accessories. For example, camera tripods or, the belt buckle
delivered with the device.
2.3 MAGNETIC
The magnet integrated at the back of the HP32 enables easy installation on metallic surfaces
such as, for example, ventilation shafts.
The magnetic fastening will not work with an accessory mounted on the tripod socket.
2.4 LANYARD
Delivered with the HP32, you will nd a lanyard that can be tightened on the HP32 for easy
hand carrying and manipulation.
3 GENERAL OPERATION
3.1 BATTERY, STAND-BY AND AUTO SWITCH OFF
The HP32 has an integrated battery. Simply connect the device to an USB port with the deliv-
ered cable to charge it.
The device has a stand-by and an auto switch off feature. After 5 minutes without user interac-
tion, the display is dimmed. Moreover, after a total of 10 minutes without user interaction, the
device is switched off to save battery.
These features do not affect data logging operations. While data logging is active,
although the screen is dimmed or off, data points are still logged and saved.
3.2 BUTTONS
3.3 DISPLAY AND HOME SCREEN
4 GENERAL SETTINGS
4.1 LANGUAGE
To change the display language, go into:
Main menu [OK] Settings Language
Available languages: English, Deutsch, Français, Italiano
4.2 DATE/TIME
To set up the date and time, go into:
Main menu [OK] Settings Date / Time
The following settings can also be set beside the date and time.
Date format: DIN, US, ISO
Time format: 12 h, 24 h
The date and time must be set correctly for proper operation of the device.
4.3 UNITS
Set units used and displayed by the HP32 into:
Main menu [OK] Settings Units
Your options are as follow:
Distance: m, ft
Temperature: °C, °F
Pressure: hPa, mmHg
Weight: kg, lbs
[POWER] Switches the HP32 on or off.
Probe connector
Tripod socket
Reset button
Integrated magnet
Lanyard xation
USB port
[DOWN] Cursor down
[LEFT]
[RIGHT]
Action buttons, their action are indicated by
the on-screen labels.
[UP] Cursor up
Parameter selection (home screen)
Status Bar
Time and date
Hold indicator
Log indicator
Battery status
Measured values
Action Buttons labels
[OK] Opens the main menu / submenus
Conrm entries
4.4 CALCULATED PSYCHROMETRIC PARAMETERS
Beside relative humidity and temperature, the HP32 can display 2 of the following 10 psychro-
metric parameters:
Dew / Frost point (Dp/Fp)
Wet bulb temperature (Tw)
Enthalpy (H)
Vapor concentration (Dv)
Specic humidity (Q)
Mixing ration by weight (R)
Vapor pressure (E)
Saturation vapor pressure (Ew)
Saturation vapor density (SVD)
Select the parameters that you want to log and/or see on the home screen within:
Main menu [OK] Settings Displayed calculated parameters
You can change the order of values on the home screen by [UP] on the home screen.
You can change the atmospheric pressure constant used in calculations in:
Main menu Settings Atmospheric pressure
5 SAVING DATA
5.1 HOLD/CAPTURE AND SPOT MEASUREMENTS
Freeze the displayed values on the home screen by pressing the [Hold/Capture] action button
on the home screen. Those values can then be:
Saved as a spot measurement with [Capture]
Released again with [Release]
5.2 DATA LOGGING
Program or start the automatic logging at regular interval with the [Log] action button. Dene
the storage location, the interval and, if you want, a start and stop time, then start the logging
operation by pressing the [Start] action button.
Your device will be automatically switched on at the given start time to start the logging.
The main menu is not accessible while data logging is active. Spot measurements are,
however, still possible.
You must name storage locations before use. Simply use the [Rename] action button
when selecting a storage location to assign it a name.
6 VIEWING SAVED DATA
6.1 SPOT MEASUREMENTS
Review the spot measurements by selecting their storage location within:
Main Menu [OK] Memory – Capture
6.2 LOGGED DATA
Review the logged data by selecting its storage location within:
Main Menu [OK] Memory – Log
Switch between graphic displays and MIN / MAX / AVG values of the 4 parameters with
[UP] and [DOWN].
7 ALARM
Set up upper, lower limits as well as hysteresis for every displayed parameter in:
Main Menu [OK] Alarms
Broken alarm limits are shown in red on the home screen.
You can also enable an acoustic signal for broken alarms in:
Main Menu [OK] Settings Buzzer
8 ADJUSTEMENT
Adjustments of the probe (humidity and temperature) can be performed in:
Main Menu [OK] Adjustement
Incorrectly adjusted probes will deliver wrong measurements. Please refer to our man-
ual for more information about probe adjustments.
9 COMPUTER CONNECTION
It is possible to connect the HP32 to a computer with the delivered USB service cable. Data
analysis and settings changes are performed with the HW4 software for which you received a
license key.
To download the HW4 software and learn more about its capabilities, please visit our webpage:
www.rotronic.com/hw4.
10 TECHNICAL DATA
General specications
Device type Handheld device (%RH / °C)
Compatible probes All HC2 and HC2A probes
Application range -10…60°C | 0…100%RH (non-cond.)
Storage range -10…45°C | 10…90%RH
Probe application range Depends on the probe
Accuracy Depends on the probe
Functions Spot measurements
Interval logging
MIN / MAX / AVG / MKT
Logging interval 1s…24h
Memory capacity 8x8000 data points
Psychrometric calculations Dew / Frost point (Dp /Fp)
Wet bulb temperature (Tw)
Enthalpy (H)
Vapor concentration (Dv)
Specic humidity (Q)
Mixing ratio by weight (R)
Vapor pressure (E)
Saturation vapor pressure (Ew)
Saturation vapor density (SVD)
Display refresh rate 1 s (typical)
Integrated Battery 48h – Logging every 30s, screen off
10h – Screen always on (typical)
PC Interface Micro-USB port,
HW4 compatible
Housing / Mechanics
Housing Material ABS, TPR
IP Protection IP65
Dimension 160 x 63 x 30 mm
Weight 185 g
Belt Buckle Set
Probe HC2A-S (not included)
HP32
Lanyard

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rotronic HYGROPALM HP32 and is the answer not in the manual?

Rotronic HYGROPALM HP32 Specifications

General IconGeneral
Humidity Range0 to 100 %RH
Resolution0.1 °C / 0.1 %RH
DisplayLCD
Weight200 g
InterfaceUSB
Protection classIP40
Measurement ParametersTemperature, Humidity
Temperature Accuracy±0.3 °C
Power SupplyBattery
Operating Conditions0 to 60 °C, 0 to 100 %RH