7-4
RETURN TO
This selection allows the operator to choose the height to which the hook returns
upon completion of a test. The choices are “START HT” or “SEARCH HT”.
ALARM
This selection allows the operator to set an alarm based on the time of day. The
alarm time is displayed as a 24 hour value and can be set to any time from 00:00
to 23:59. An audible “double beep” sounds at the alarm time.
SPEED
This selection allows the operator to set the Z axis speed of the test hook as
controlled by the system during testing. The speed can be adjusted from a
minimum of 4 mils per second to a maximum of 240 mils per second with a default
value of 100 mils per second. The default setting is edited by pressing the left or
right mouse buttons to increase or decrease the value, then press the center
mouse button to accept the value.
NDT HOLD
This selection allows the operator to enter a hold time for the non-destruct test
pull. The default setting is edited by pressing the left or right mouse buttons to
increase or decrease the value, then press the center mouse button to accept the
value.
SEMI AUTO
This selection allows the operator to toggle the semi automatic testing mode on or
off. Press the center mouse button to toggle the setting.
MUST FAIL CODE
This selection allows the operator to set the system to prompt for a fail code and
wait until the code is entered before resuming test mode. Press the center mouse
button to toggle the setting.
TEST STOPS AT
This selection allows the operator to set the percentage of maximum test force
below which the test is concluded. The default setting is edited by pressing the left
or right mouse buttons to increase or decrease the value, then press the center
mouse button to accept the value.
HOOK Z SPEED
This selection allows the operator to set the Z axis speed of the test hook as
controlled by the left and right mouse buttons. The speed can be adjusted from a
minimum of 1 (50 mils per second) to a maximum of 10 (500 mils per second) with
a default value of 5 (250 mils per second). The default setting is edited by
pressing the left or right mouse buttons to increase or decrease the value, then
press the center mouse button to accept the value.