E-6
Type K
Measuring range : -150°C to 1350°C (-238°F to 1999°F).
Resolution : 0.1°C , 1°C , 0.1°F , 1°F
Measurement rate : Once per second.
Basic accuracy : (@ 23±5°C calibration) accuracy are ±
(…% of reading + degree) at 18°C to
28°C with relative humidity up 80%.
Function Resolution Range Accuracy
-150°C~0°C ±( 0.2%rdg + 1.0°C)
0.1°C
0°C~200°C ±( 0.1%rdg + 1.0°C)
°C
1°C
200°C~1350°C ±( 0.2%rdg + 2°C)
-238°F~32°F ±( 0.2%rdg + 2°F)
0.1°F
32°F~200°F ±( 0.1%rdg + 2°F)
° F
1°F
200°F~1999°F ±( 0.2%rdg + 4°F)
Temperature Coefficient:
0.1 times the applicable accuracy specification per °C from 0°C to
18°C and 28°C to 50°C ( 32°F to 64°F and 82°F to 122°F ).
Note :
The basic accuracy specification does not include the
error of the probe. Please refer to the probe accuracy
specification for additional details.