EasyManua.ls Logo

RS PRO 1327K - Setting Mode; Parameter Settings for Measuring

RS PRO 1327K
19 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
E-15
e Reading the memory data: Press and hold the R button for
3 seconds to enter READ mode. Press the ▲ or ▼ buttons
to select to desired memory data location. Press the
MODE
button to exit READ mode.
Ⅷ. SETTING MODE
Pull trigger to turn the meter on. Press & hold the MODE button
for 3 seconds until “SET” appears on the LCD.
1. Parameter settings for measuring:
Under the infrared parameter setting mode, the LCD will
show . Press the
MODE button to select setting of “°C / °F”,
“ALM ON/OFF”, “ÈALM” “ALMÇ” and emissivity “
ε
”. Pull the
trigger to exit setting mode and return to normal measurement
mode.
2. Parameter settings:
c “°C / °F”: Temperature unit / °F, press ▲
or ▼ to select units
°C or °F.
d ALM (Alarm function ON/OFF): Press ▲ or ▼ to select alarm
function on or off.
e ÈALM (Lo Alarm Function): Press ▲
or ▼ to set a value
as an alarm value. When this Lo alarm value is exceeded,
the beeper will sound and the “ÈALM” symbol will appear
on the LCD.
f ALMÇ (Hi Alarm Function): Press ▲
or ▼ to set up a value
as an alarm value. When this Hi alarm value is exceeded,
the beeper will sound and the “ALMÇ” symbol will appear
on the LCD.
Example:
To sound the alarm if the temperature exceeds 100°C or
falls below 0°C, set the Hi Alarm point to100°C and the
Lo Alarm point to 0°C.
SET