This product is medical device.
Make sure to read this user manual before using.
This user manual is to assure proper installation and use.
Pay attention to read this user manual in order to making the best use of this product and to assure the prolonged lifespan before using
Specially, pay attention the contents with marks such as , and .
<User>
• Qualified Professional
<Intended use>
• Surgical bone cutting, shaving, polishing, grinding, drilling.
<Classification of equipment>
• Type of protection against electric shock: ClassⅠ equipment
• Degree of protection against electric shock: Type BF applied part
• Classification according to the degree of protection against ingress of water as detailed in the current edition of IEC 60529
- Control Box: IPX0
- Foot Controller: IPX8
• Not suitable product in the atmosphere that exist flammable anesthetic mixture with air or with oxygen.
• User sterile product (non-sterile product at shipment)
• Classification according to mode of operation: Intermittent operation
• Applied parts: DRILL bur, SAW blade, SHAVER blade
<Use time>
• Loading time: 3 minutes
• Resting time: 10 minutes
INTRODUCTION
2
Operating condition
Transport and storage condition
Temperature (℃)
Relative humidity (%)
Atmospheric pressure (hPa)
Temperature (℃)
Relative humidity (%)
Atmospheric pressure (hPa)
10~30
10~80
700~1,060
-20~60
10~90
500~1,060
<Operation & Storage condition>