CAUTION
Pay attention to use considering a patient’s safety first.
Check the working condition to see if there is any problem before operation.
Stop working and verify the product if there is any signs of vibration, calescence or strange noise before or during operation.
Contact the manufacturer if the problem is abnormal or difficult to solve.
Prepare extra consumables before using for safety. In case blood or saline solution flows into the head of the angle, which may cause malfunctioning.
Change Bur, Blade after the motor stops completely. Otherwise, Inside of angle could be damaged by the Bur, Blade, and it will cause an accident.
Please sterilize and lubricate products immediately after use following the manual. Otherwise, the product can be damaged due to coagulation of blood
or remains from operation.
The control box and foot controller are not steam sterilization objects.
The input power is available at AC 100-120 V / 220-240 V.
(Check input voltage before connecting the AC power.)
Dispose the irrigation tube after use as medical waste.
Contact to manufacturer if it shows the fault display on the LCD. It may cause an accident.
Do not drop from high or give a shock on the product. It can damage the product.
Do not give any damage to irrigation tube, it will not work normally due to the tube is broken or taken off.
Do not insert oil into the motor. It will damage bearing or cause calescence.
This product is designed to be used indoors. Do not use outdoors.
Do not place or use the product in humid or dusty places.
Use products and consumables approved by manufacturer. Otherwise there might be a risk of malfunction or accident.
Replace the motor cord immediately when it is worn out or damaged to prevent electric shock for user or patient.
Product specification and design can be changed for improvement of the product.
User is responsible for operation and maintenance of the product.
The company did not sterilize the product. Please be sure to sterilize before operation. (User sterilization)
US Federal law restricts this device to sale by or on the order of a dentist.
USER GUIDELINES
4