Quick Reference Guide
Tissue-Tek® DRS™ 2000
QRG | DRS 2000 | 2011 | V-3 page 3 | 3
Maintenance
Daily
Exterior and interior surfaces:
1. Clean the plastic and painted areas by
wiping with a clean cloth (use a glass
cleaner)
2. Clean the interior of the solution area by
wiping with a clean cloth
Daily
Washing stations and drain:
1. Check wasking stations, clean the drain
using bleach
Note: Do not clean the whole sink area with bleach.
This can cause corrosion of the stainless steel
Weekly
Solution reservoirs:
1. Solution Reservoirs should be washed with
tap water and a mixture of mild detergent
and bleach
2. When stains and residue are difficult to
Remove, use bleach. Afterwards clean the
reservoirs with water and let them dry
(<60˚ C).
Weekly
Baskets:
1. Clean slide baskets with xylene to remove
residues of film/glue
2. Rinse with alcohol and air dry
Bi-monthly
Replace the activated carbon filters
Monthly
Drying stations:
1. Remove all inner components of the drying
stations
2. Remove the residue and any debris from the
components with water and soap and air-
drying
3. Place the components back
Yearly
Preventive maintenance by a Sakura representative
Or Sakura service engineer.
Clean the solution reservoirs
Replace the activated carbon
filters