EasyManuals Logo

Salter Labs PRO2 Check User Manual

Salter Labs PRO2 Check
20 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #18 background imageLoading...
Page #18 background image
PRO
2
check
®
Ultrasonic Oxygen Indicator
VERIFICATION & CALIBRATION INSTRUCTIONS
PURE OXYGEN VERIFICATION PROCEDURE
To quickly verify that the indicator unit is accurately measuring the
concentration of oxygen being produced by a concentrator, perform the
following steps:
1.
Connect the inlet port on the PRO
2
check
®
unit to a USP oxygen
source
(cylinder or liquid oxygen) via a supply tube.
2.
Turn the O
2
delivery flow to 2 LPM.
3.
Press the on/offbutton to turn unit on.
4.
When unit stabilizes the display will read - "Concentrator O
2
91.8"
(approx. concentration.).
NOTE: When in the Concentrator Mode, the electronic components of the
PRO
2
check
®
indicator automatically compensate for the argon found in
the gas produced by an oxygen concentrator. As a result the display will
read 90% 2%) unless this function is over-ridden by pushing the
"pure O
2
/calibrate" button).
5.
Press and continue to hold down the pure O
2
/calibrate button.
6.
Unit will display Pure O
2
and a numeric reading.
7.
If the unit is properly calibrated, the display should read between
99% and 101%. The
PRO
2
check is ready to use.
8.
If the unit display reads "err, "---", "-U-" or any other reading
other than 99.9 2%), the unit should be recalibrated (See below).
QUICK “ONE STEP ” P URE OXYGEN CALIB RATIO N
The following procedure can be used to quickly and easily calibrate the
Ultrasonic Oxygen Concentrator Indicator to pure oxygen:
1.
Follow steps 1 through 6 of "Pure Oxygen Verification Procedure"
above.
2.
When the unit displays "Pure O
2
" and a numeric reading, continue
holding down the pure O
2
/calibratebutton.
3.
After a few seconds, the LCD display (while continuing to read
"Pure O
2
") will start to alternately display "CAL" and number.
4.
Continue to hold the pure O
2
/calibratebutton down until
the display stops alternating and simply displays “Pure O
2
100”.
5.
At that point, the unit is calibrated to pure oxygen and ready for use.
NOTE: If unit displays an O
2
concentration reading outside of the ± 2%
tolerance range (eg. < 98% or >102%) unit will not “Quick Cal” and the
Full Two Step procedure on the following page must be followed.
6

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Salter Labs PRO2 Check and is the answer not in the manual?

Salter Labs PRO2 Check Specifications

General IconGeneral
BrandSalter Labs
ModelPRO2 Check
CategoryMeasuring Instruments
LanguageEnglish