Salter Labs PRO2 Check - Spot Checking Oxygen Gas Cylinders

Salter Labs PRO2 Check
20 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
PRO
2
check
®
Ultrasonic Oxygen Indicator
2
INSTRUCTIONS FOR SPOT CHECKING
LIQUID OXYGEN VESSELS or OXYGEN CYLINDERS
Although the PRO
2
check
®
Ultrasonic Oxygen Indicator is designed, as its
name implies, to check the various functions of an oxygen concentrator, it
can also be used to perform spot checks 2%) of gaseous or liquid
oxygen.
To utilize the PRO
2
check
®
device for this purpose follow this procedure:
1.
Connect the inlet port on the PRO
2
check
®
unit to a gaseous or
liquid
oxygen source (cylinder or liquid oxygen) via a supply tube.
2.
Turn the oxygen flow on to 2 LPM.
3.
Press the on/offbutton to turn the PRO
2
check
®
unit on.
4.
When unit stabilizes the unit display will read "Concentrator O
2
91.8" (approx concentration.)
5.
Press and continue to hold down the pure O /calibratebutton.
6.
Unit will display "Pure O2" and a numeric reading.
7.
If the source gas is oxygen, the display should read between 99%
and 101%.
8.
If the unit display reads "err", "---", "-U-" or any other reading
other than 99.9 2%), the source gas may not be pure oxygen and
should be quarantined for retesting using a more sophisticated method
such as a Servomex* oxygen analyzer.
NOTE: PRO
2
check
®
Ultrasonic Oxygen Indicator can be used to spot
check
the identity and purity of gaseous or liquid oxygen only, it cannot be
used to
perform or verify U.S.P. oxygen (> 99.0%) testing since it cannot
provide
the required U.S.P. oxygen (> 99.0%) accuracy of ± 0.1%.
* Servomex Inc. - Sugarland, TX