Control panel and keypad
1. LCD panel
2. Upper limit regulator:
This regulator is used to set the upper
temperature limit.
3. OVERHEAT indicator:
This indicator lights when the chamber
temperature reaches the upper limit.
4. MENU Key:
Press this key to access the menu.
5. Contrast knob:
Turn this knob to adjust the contrast of the
LCD.
6. BUZZER Key:
Press this key to silence the buzzer.
Note:
It is not possible to silence the
buzzer for an upper limit temperature
alarm.
7. CE Key:
Press this key to clear the entered value
when entering a setting.
8. Cursor Keys (Up, Down, Left, and Right):
Use these keys to move the cursor on the
LCD panel.
9. ENTER Key:
Press this key to select a menu command.
When entering a setting, pressing this key to
move to the next parameter.
10. Numeric Keys
11. H
2
O
2
Key:
Press this key to start H
2
O
2
decontamination.
Refer to page 40 for details.
The following optional components must be installed to perform decontamination.
UV Lamp Add-on Kit (MCO-19UVS) (This is provided as standard equipment for the MCO-19AIC(UV).)
H
2
O
2
Decontamination Kit (MCO-HL)
H
2
O
2
Generator (MCO-HP)
1
2
4
5
6
7
8
9
10
11
3