Questions and Answers

  • J
    Jennifer ChandlerAug 17, 2025
    Why is the CO consumption too high in my Sanyo MCO-18AC UV incubator?
    • S
      Sara HayesAug 17, 2025
      If the CO consumption is too high, consider the following possible causes: * The door is being opened too often. * There is a gas leak at a connection or pinhole in the tube. * The inner door gasket isn't completely sealed. * The access port in the upper left corner is open.
  • T
    Trevor LucasAug 21, 2025
    What to do if chamber temperature is not equal to the set temperature in Sanyo Accessories?
    • E
      Ethan NovakAug 21, 2025
      If the chamber temperature isn't reaching the set temperature, ensure that the ambient temperature is always at least 5oC lower than the set temperature and that the outer door is closed while the inner door isn't left open.
  • L
    Lori LongAug 25, 2025
    What to do if the CO gas density does not coincide with the set value in Sanyo MCO-18AC UV Accessories?
    • B
      Brianna CarlsonAug 26, 2025
      If the CO gas density doesn't match the set value, ensure that the secondary pressure is set to 0.03MPaG (0.3kgf/cm2G, 4.3psiG), the gas tube isn't clogged or kinked, and the humidifying pan is filled with sterile distilled water.
  • B
    brandy82Aug 28, 2025
    Why normal cultivation cannot be done and chamber gas density is suspect in Sanyo MCO-18AC UV Accessories?
    • J
      Jordan MartinAug 29, 2025
      If normal cultivation is not possible and chamber gas density is suspect, it may be because the environment around the incubator is not normal, there is a source of contaminated gas nearby, or the incubator is installed in an enclosed space.
  • P
    Patricia JordanSep 1, 2025
    How to troubleshoot a Sanyo Accessories incubator that does not operate at all?
    • L
      Luis GreenSep 1, 2025
      If the incubator isn't operating, make sure it's properly plugged into a power outlet and that the circuit breaker at the power source is on. Also, check for any power failures.
  • C
    Charlotte MillerSep 5, 2025
    What to do if the chamber humidity does not rise in Sanyo MCO-18AC UV?
    • L
      Leslie NunezSep 5, 2025
      If the chamber humidity isn't rising, make sure the humidifying pan is filled with sterile distilled water.
  • R
    Ryan GraySep 8, 2025
    Why is the key operation disabled on my Sanyo Accessories?
    • Y
      ysmithSep 8, 2025
      If the key operation is disabled, the key lock function might be activated. Check if the key lock function is set to ON mode.
  • H
    Holly EvansNov 13, 2025
    What to do if the alarm function and buzzer operate on my Sanyo MCO-18AC UV Accessories incubator?
    • J
      Jodi SmithNov 13, 2025
      If the alarm and buzzer are activated, it could be due to several reasons: * The chamber temperature isn't matching the set value. * The chamber CO density is not at the desired level. * The secondary pressure of the pressure regulator isn't correct (0.03MPaG, 0.3kgf/cm2G, 4.3psiG). * The tube connecting the pressure regulator and incubator isn't secure. * The upper temperature limit is less than 1oC above the set chamber temperature. * The set temperature has been changed, or the door was open for too long. * A low-temperature item was placed inside. * The gas tube has come loose, or there's a gas leak. * The set gas density was altered. * The gas cylinder is empty; check the CO cylinder's primary pressure weekly. A primary pressure below 2 3.8MPaG (38kgf/cm2G) indica...

Summary

Introduction

Precautions for Safe Operation

Environmental Conditions

Incubator Components

Control Panel and Keypad

Remote Alarm Terminal

Installation Site

Installation

Connection of CO2 Gas Cylinder

Preventing Contamination

Precautions for Cultures

Start-up of Incubator

Setting Chamber Temperature and CO2 Density

UV Lamp

Precautions When Using the UV Lamp

Provides essential safety guidelines for operating the UV lamp, including avoiding direct eye exposure and using protective covers.

Setting the UV Lamp ON Period

Details the procedure to adjust the duration the UV lamp remains active after the outer door is closed.

Lighting the UV Lamp for 24 Hours

Explains how to set the incubator for continuous UV lamp operation for chamber decontamination purposes.

Key Lock Function

Water Level Sensor

Setting of Alarm Resume Time

Operation After Power Failure

Describes how the incubator resumes operation with previous settings after a power interruption.

Routine Maintenance

Cleaning the Chamber and Inner Attachments

Step-by-step guide for cleaning and sterilizing the incubator's interior components and chamber.

Filling the Humidifying Pan

Calibration

Temperature Calibration

Explains the procedure to calibrate the incubator's temperature sensor for accurate readings.

CO2 Calibration

Details the process for calibrating the CO2 sensor to ensure accurate gas concentration measurements.

Disposing of the CO2 Incubator

Automatic CO2 Cylinder Changeover

Stacking Incubators

Safety Check Sheet

Sanyo MCO-18AC UV Specifications

General IconGeneral
BrandSanyo
ModelMCO-18AC UV
CategoryAccessories
LanguageEnglish

Related product manuals