151
60
3. Input the CO
2
density of standard gas to be used, press the MENU Key to display the menu, select OK,
and press the ENTER Key.
Ɣ Calibration method for CO
2
gas density
1. Select Standard Gas Calibration from the Select Tools (2/2) Display, press the MENU Key to display
the menu, select OK, and press the ENTER Key.
2. The system check will start.
S e l e c t T o o l s ( 2 / 2 )
S t a n d a r d G a s S e t t i n g
S t a n d a r d G a s C a l i b r a t i o n
M E N U
O K
S v c
C a n c e l
1
S t a n d a r d G a s S e t t i n g
3
C
O 2 S T D G a s D e n s i t y 5 . 0 %%
4
5
6
M E N U
O K
C a n c e l
1
S t a n d a r d G a s C a l i b r a t i o n S t e p 1
3
S t d G a s C a l i b S y s t e m C h e c k ࡮࡮࡮࡮
4
5
CO 2 0 . 0 %
STG 5 . 0 %
6