CALIBRATION
Temperature/CO
2
/O
2
calibration
1. Press the MENU Key from the Top Display to display the menu, select Tools, and press the ENTER
Key.
2. Select Temp/CO2/O2 Calibration from the Select Tools Display, press the MENU Key to display the
menu, select OK, and press the ENTER Key.

3. The Temp/CO2/O2 Calibration Display will appear.
Example
Enter the value shown below if the displayed chamber temperature is 37.0
o
C but the actual temperature
is 36.5
o
C.
Temperature span: Enter 365.
S e l e c t T o o l s ( 1 / 2 )
T e m p / C O 2 / O 2 C a l i b r a t i o n
A l a r m Se t t i n g
L C D / D A Q S e t t i n g
U V S e t t i n g
K e y L o ck P W S etting
D a t e T i m e
M E N U
O K
S v c
C a n c e l
1
T e m p / C O 2 / O 2 C a l i b r a t i o n
3
P V
Volt
T e m p S p a n 3 6 . 5
o
C37 . 0
o
C 0 . 5°C
4
C O 2 Z e r o 3 . 9 1 V 5 0 0
5
C O 2 S p a n 0 . 0 % 5 . 0 % 3 . 1 6 V 7 0 0
O 2 Z er o 5 0 0
6
O 2 S p a n 0 . 0 % 5 . 0 % 5 8 m V 5 0 0
.
1
T e m p 3 7 . 0
o
C C O 2 5 . 0 %
O 2 5 . 0 %
2
3
4
5
S t a t u s
6
O K
7
37.
.
.
M E N U
S e t
L o g
T o o l s