EasyManua.ls Logo

Sanyo MDF-U333 - Function Mode Settings

Sanyo MDF-U333
53 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
29
10.
10.10.
10.Function mode
Functions are as follow.
F00 To return to PV display
F01 To change temperature of high temperature alarm
F02 To change temperature of low temperature alarm
F03 To change temperature of high temperature alarm
(Applied for lower compartment of MDF-U537D)
F04 To change temperature of low temperature alarm
(Applied for lower compartment of MDF-U537D)
F05 To set compressor delay start time
F06 To input service code (384)
F07 To adjust zero for temperature sensor
(Upper compartment sensor for MDF-U537D)
F08 To adjust zero for temperature sensor
(Not available for MDF-U333)
(Solenoid valve control sensor for MDF-U537)
(Lower compartment sensor for MDF-U537D)
F09 Not available
F10 Not available
F11 Not functioned. Return to PV display.
F12 To display temperature of temperature sensor (per 0.1)
F13 To display temperature of MDF-U537/U537D compressor protect sensor (per 1)
F14 To display temperature of temperature sensor (per 0.1)
(Solenoid valve control sensor for MDF-U537)
(Lower compartment sensor for MDF-U537D)
F15 Not available
F16 Not functioned. Return to PV display.
F17 To set or check model code
model code 001 MDF-U333
model code 002 MDF-U537
model code 003 MDF-U537D
F18 Preset value for second offset performance: -26 (for MDF-U537/U537D)
F19 Not functioned. Return to PV display.
F20 Not functioned. Return to PV display.
F21 To set communication ID000255
F22 To set communication mode
F23,24,2629 With input each function code, press SET key to sound buzzer 1 second
continuously to maintain input status.
F25 To set alarm ring back time
Setting value: 000,010,020,030,040,050,060
Setting method
In PV display, keep key pressing over 5 seconds to enter function mode and will be
displayed “F00”. Press key to input required function code then press
SET key again.
F00 Not functioned. In F00 displayed, press SET key to return PV display.