MDF-U333
Temperature uniformity and fluctuation
Test condition
Mains : 1φ100V50/60Hz
AT : 35
SV -25℃・-20℃・-30
No load Measure with type T sensor
Measuring pointCenter point of each container
(1).SV=-25℃
Temperature(
Fluctuation(
Top -21.8
-23.2
Middle -24.3 ± 2.0
-24.6
Bottom -24.2
Uniformity Max./Min -21.8/-24.6
Range(deg) 2.8(±1.4)
(2).SV=-20℃
Temperature(
Fluctuation(
Top -17.1
-18.3
Middle -19.2.6
-19.4
Bottom -19.0
Uniformity Max./Min -17.1/-19.4
Range(deg) 2.3(±1.15)
(3).SV=-30℃
Temperature(
Fluctuation(
Top -26.3
-28.1
Middle -29.3 ±1.7
-29.4
Bottom -29.1
Uniformity Max./Min -26.3/-29.4
Range(deg) 3.2(±1.6)
noteThese data are reference only.
Measuring point
1
2
3( 1/2 center air)
4
5
Measuring point
Measuring point
1
2
3( 1/2 center air)
4
5
5
1
2
3( 1/2 center air)
4
42−