32
F17 Model code setting or checking
001 MDF-U333
002 MDF-U537
002 MDF-U537D
Followings are initial setting values
Internal temp. SV -30 (for U537D : both compartments)
Key lock mode 0 OFF
High temperature alarm SV+10(for U537D : both compartments)
Low temperature alarm SV-10(for U537D : both compartments)
0 adjustment+0℃(for temperature sensor
Compressor start delay time 1 minute
Communication ID 000
Communication mode 000
Alarm ring back time 030 (30 minutes)
Set model code as MDF-U537, MDF-U537D to activate compressor protect function.
F21 To set serial communication ID.
Input range: 000255 000 means that communication is OFF
F22 To set serial communication mode.
Control mode 3
rd
digit 0 Local initial value
1 Remote
Baud rate 2
nd
digit 0 2400bps initial value
1 4800bps
2 9600bps
1
st
digit is not used.
Note Set control mode as “remote” unable to change SV and to activate defrosting.
F25 To set value for alarm recovery (ring back) time.
Input value: 000,010,020,030,040,050,060 (In 000, recovery function is OFF)
11.
11.11.
11.Differential COMP. ON/OFF point on program
COMP ON permit Activate at +0.0 higher than SV
COMP OFF Activate at -0.6 lower than SV