EasyManuals Logo

Sanyo MDF-U53VC User Manual

Sanyo MDF-U53VC
68 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #25 background imageLoading...
Page #25 background image
22
F31: <Purpose> Buzzer setting during filter alarm
<Operation> Input F31 and press SET key to display “001” (initial).
Change to the desired value with key and key.
Press SET key to revert to PV display.
000: Buzzer off
001: Buzzer on
F32: <Purpose> To display accumulation time of condensing fan motor
<Operation> Input F32 and press SET key to display F32 and ‘xx.x’ (accumulation
time) alternately. Press SET key to revert to PV display.
8. Differential (The point of compressor ON and OFF )
COMP ON:
COMP OFF:
SV +0.4
SV -2.2 (for SV is -90~-60), SV -0.5 (for SV is -59~-50)
9. Offset value
1) PV+0.5 is the offset value to adjust the difference between temperature of temp. sensor
and the center of internal chamber.
2) PVat + (-3.0C) is the offset value to correct the ambient temperature.
*Note) PVat = Temperature of AT sensor
10. Remote alarm terminal
In normal condition:
In alarm condition &
power failure :
Remote alarm contact is N.O. N.C.
Remote alarm contact is N.C. N.O.
11. Cycling of High stage compressor
If the filter sensor is higher than or equal to 60, the High stage compressor will be turned off.
When the filter sensor is within 10 of the ambient temperature sensor, the High stage
compressor will be allowed to turn back on.
12. Sensor failure
(1) Temp. sensor
Open circuit:
E01 and 50 are displayed alternately, the buzzer beeps intermittently
and remote alarm contact outputs.
The compressor would be allowed to turn on.
Press BUZZER key to stop the buzzer beeping.
Short circuit:
E02 and -170>PV>-180 are displayed alternately, the buzzer beeps
intermittently and remote alarm contact outputs.
The compressor would be allowed to turn on.
Press BUZZER key to stop the buzzer beeping.
(2) Cascade sensor
Open circuited: E03 and PV are displayed alternately, the buzzer beeps intermittently
and remote alarm contact outputs.
The resistance value would be 㺙㱅 and temperature would be -34.1
or lower. Compressor (L) is not forcibly turned off.
Press BUZZER key to stop the buzzer beeping.
Short circuited: E04 and PV are displayed alternately, the buzzer beeps intermittently
and remote alarm contact outputs.
Press BUZZER key to stop the buzzer beeping.
The resistance value would be 0 and temperature would be -12 or
higher that causes to be compressor (L) failure. Compressor (L) is
forcibly turned off.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Sanyo MDF-U53VC and is the answer not in the manual?

Sanyo MDF-U53VC Specifications

General IconGeneral
BrandSanyo
ModelMDF-U53VC
CategoryFreezer
LanguageEnglish

Related product manuals