Specifications of sensor
The following shows the temperature in thermal sensor (502AT-1) and its resistance value.
Temp.
(C)
Resistance
Value (kΩ)
Temp.
(C)
Resistance
Value (kΩ)
Temp.
(C)
Resistance
Value (kΩ)
Temp.
(C)
Resistance
Value (kΩ)
50
154.5
36
71.80
22
35.65 0 13.29
49
145.9
35
68.15
21
33.99 5 10.80
48
137.8
34
64.71
20
32.43 10 8.84
47
130.2
33
61.48
19
30.92 15 7.20
46
123.1
32
58.43
18
29.50 20 6.01
45
116.5
31
55.55
17
28.14 25 5.00
44
110.2
30
52.84
16
26.87 30 4.17
43
104.4
29
50.23
15
25.65 35 3.50
42
98.87
28
47.77
14
24.51 40 2.96
41
93.70
27
45.45
13
23.42 45 2.51
40
88.85
26
43.26
12
22.39 50 2.13
39
84.18
25
41.19
11
21.41 55 1.82
38
79.80
24
39.24
10
20.48 60 1.56
37
75.67
23
37.39
5
16.43 65 1.35
The following shows the temperature in thermal sensor (PT1000ȍ) and its resistance value.
Temp.
(C)
Resistance
Value (Ω)
Temp.
(C)
Resistance
Value (Ω)
Temp.
(C)
Resistance
Value (Ω)
140
450.83
70
729.99 0 1000.0
130
491.47
60
769.02 10 1038.0
120
531.83
50
807.87 20 1076.0
110
571.92
40
846.58 30 1113.8
100
611.76
30
885.13 40 1151.4
90
651.38
20
923.55 50 1189.0
80
690.78
10
961.84 60 1226.4