EasyManua.ls Logo

Sartorius ACCULAB ATILON ATL-84 - User Manual

Sartorius ACCULAB ATILON ATL-84
59 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
Standard Unit With Built in Calibration
Factory Verified and with Built in
Calibration
ALT124 £936.85 ALT124-I £1,088.85 ALT124-V £1,157.25
ALT84 £838.05 ALT84-I £990.05 ALT84-V £1,058.45
ALT224 £1,012.85 ALT224-I £1,164.85 ALT224-V £1,233.25
ALT153 £526.45 ALT153-I £678.45 ALT153-V £746.85
ALT423 £625.25 ALT423-I £777.25 ALT423-V £845.65
ALT623 £716.45 ALT623-I £868.45 ALT623-V £936.85
ALT822 £572.05 ALT822-I £724.05 ALT822-V £792.45
ALT2202 £625.25 ALT2202-I £777.25 ALT2202-V £845.65
ALT4202 £701.25 ALT4202-I £853.25 ALT4202-V £921.65
ALT6202 £792.45 ALT6202-I £944.45 ALT6202-V £1,012.85
ALT6201 £663.25 ALT6201-I £815.25 ALT6201-V £883.65
ALT8201 £708.85 ALT8201-I £860.85 ALT8201-V £929.25
Question and Answer IconNeed help?

Do you have a question about the Sartorius ACCULAB ATILON ATL-84 and is the answer not in the manual?

Overview

The Acculab ATILON series comprises 11 models of analytical and precision balances, designed for high-end technology and low-end pricing. These balances are suitable for various applications, including those requiring legal metrology. They offer a comprehensive variety of features and outstanding performance, making them a versatile tool for laboratories and other settings.

Function Description

The ATILON balances are primarily used for precise weighing. They offer several application programs to enhance functionality:

  • Counting: Determines the number of parts that have approximately equal weight. Users can set a reference sample quantity and the balance calculates the total number of parts.
  • Weighing in Percent: Allows users to obtain weight readouts in percent, relative to a reference weight.
  • Calculation: Enables calculation of weight values using a multiplier or divisor. This can be used for determining weight per unit area or "gsm" weight.
  • Animal Weighing/Averaging: Designed to determine the weights of unstable samples (e.g., live animals) or to determine weights under unstable ambient conditions. The balance calculates the average of a defined number of individual weighing operations.
  • Net-total Formulation: Allows users to weigh individual components and obtain both the total weight and the individual weights of the components. Up to 99 components can be stored.
  • Totalizing: Adds successive, mutually independent weight values to a total that exceeds the capacity of the balance.
  • Density Determination: Determines the density of solid substances using the buoyancy method. Results can be displayed with one or no decimal place.
  • Mass Unit Conversion: Changes the weight value displayed from the basic weight unit to any of 4 application weight units.

The balances also feature a built-in, motorized calibration at the touch of a key, available on all models, with standard or verified models for legal metrology.

Important Technical Specifications

The ATILON series includes various models with different weighing capacities, readabilities, and other specifications. Here are some key technical specifications across different models:

General Specifications (for all ATL...-I and ATL...-V models):

  • Built-in motorized calibration weight: Yes
  • AC power source/power requirements: 230 V or 115 25 V, +15% to –20%, 48 – 60 Hz
  • Power consumption: maximum 16 VA; typical 8 VA (STNG6)
  • Approx. hours of operation with the YR805Z rechargeable battery pack (backlighting on): 35 hours

Specific Model Specifications (examples):

  • ATL-224, ATL-224-I:

    • Weighing capacity: 220 g
    • Readability: 0.0001 g
    • Tare range (subtractive): 220 g
    • Repeatability (std. deviation): ≤±g 0.0001
    • Linearity: ≤±g 0.0002
    • Response time (average): 2.5 s
    • Sensitivity drift within +10 to +30°C: ≤±/K 2 · 10⁻⁶
    • External calibration weight: 200 g (E2)
    • Operating temperature range: +10 ... +30 °C (273 ... 303 K, 50° ... 86 °F)
    • Net weight, approx.: 4.4 | 4.7 kg
    • Weighing pan size: 90 mm Ø
    • Dimensions (W × D × H): 230 × 303 × 330 mm
  • ATL-623, ATL-623-I, ATL-423, ATL-423-I, ATL-153, ATL-153-I:

    • Weighing capacity: 620 g (ATL-623), 420 g (ATL-423), 150 g (ATL-153)
    • Readability: 0.001 g
    • Tare range (subtractive): 620 g (ATL-623), 420 g (ATL-423), 150 g (ATL-153)
    • Repeatability (std. deviation): ≤±g 0.001
    • Linearity: ≤±g 0.002 (ATL-623, ATL-423), 0.003 (ATL-153)
    • Response time (average): 1 s (ATL-623, ATL-423), 1.3 s (ATL-153)
    • Sensitivity drift within +10 to +30°C: ≤±/K 2 · 10⁻⁶ (ATL-623, ATL-423), 5 · 10⁻⁶ (ATL-153)
    • External calibration weight: 500 g (E2) (ATL-623), 200 g (F1) (ATL-423), 100 g (F1) (ATL-153)
    • Net weight, approx.: 3.2 | 3.6 kg (ATL-623), 3.2 | 3.6 kg (ATL-423), 2.6 | 3.0 kg (ATL-153)
    • Weighing pan size: 115 mm Ø
    • Dimensions (W × D × H): 230 × 303 × 136 mm
  • ATL-6202, ATL-6202-I, ATL-4202, ATL-4202-I, ATL-2202, ATL-2202-I, ATL-822, ATL-822-I:

    • Weighing capacity: 6200 g (ATL-6202), 4200 g (ATL-4202), 2200 g (ATL-2202), 820 g (ATL-822)
    • Readability: 0.01 g (ATL-6202, ATL-4202, ATL-2202), 0.001 g (ATL-822)
    • Tare range (subtractive): 6200 g (ATL-6202), 4200 g (ATL-4202), 2200 g (ATL-2202), 820 g (ATL-822)
    • Repeatability (std. deviation): ≤±g 0.01 (ATL-6202, ATL-4202, ATL-2202), 0.001 (ATL-822)
    • Linearity: ≤±g 0.02 (ATL-6202, ATL-4202, ATL-2202), 0.002 (ATL-822)
    • Response time (average): 1.1 s (ATL-6202, ATL-4202, ATL-2202), 1.0 s (ATL-822)
    • Sensitivity drift within +10 to +30°C: ≤±/K 2 · 10⁻⁶ (ATL-6202, ATL-4202, ATL-2202), 5 · 10⁻⁶ (ATL-822)
    • External calibration weight: 5000 g (E2) (ATL-6202), 2000 g (F2) (ATL-4202), 2000 g (F1) (ATL-2202), 500 g (F2) (ATL-822)
    • Net weight, approx.: 3.1 | 3.5 kg (ATL-6202, ATL-4202, ATL-2202), 2.0 | 2.6 kg (ATL-822)
    • Weighing pan size: 180×180 mm (ATL-6202, ATL-4202, ATL-2202), 115 mm Ø (ATL-822)
    • Dimensions (W × D × H): 230×303×91 mm (ATL-6202, ATL-4202, ATL-2202), 230×303×87 mm (ATL-822)
  • ATL-8201, ATL-8201-I, ATL-6201, ATL-6201-I:

    • Weighing capacity: 8200 g (ATL-8201), 6200 g (ATL-6201)
    • Readability: 0.1 g
    • Tare range (subtractive): 8200 g (ATL-8201), 6200 g (ATL-6201)
    • Repeatability (std. deviation): ≤±g 0.1
    • Linearity: ≤±g 0.3 | 0.1
    • Response time (average): 1 s
    • Sensitivity drift within +10 to +30°C: ≤±/K 1.0 · 10⁻⁶
    • External calibration weight: 5000 g (F2)
    • Net weight, approx.: 2.7 | 3.5 kg
    • Weighing pan size: 180×180 mm
    • Dimensions (W × D × H): 230×303×91 mm

Usage Features

The ATILON balances are designed for user-friendliness and versatility:

  • Large, easy-to-read, backlit display: Ensures clear visibility of readings.
  • Front level indicator: For easy checking of the leveling status.
  • ISO/GLP printouts: Compliant printouts with suitable printer for documentation.
  • Large, frameless draft shield: With 3 sliding doors for simple and convenient weighing on all 0.1-mg models.
  • Sturdy ABS housing: With easy-to-clean surfaces.
  • Digital filters: 4 digital filters for adaptation to ambient conditions, ensuring the fastest weighing results along with the highest accuracy.
  • RS-232C interface port: For connecting to peripheral devices like printers or computers.
  • Option for below-balance weighing: With built-in hanger.
  • Anti-theft locking device: With accessory LC-1.
  • Warmup Time: Models ATL-153 and ATL-822 require 2 hours, while all other models require 30 minutes for accurate results.
  • Operation Outside the Temperature Range: The balance can operate beyond the temperature range of +10...+30°C (50°...86°F), but specifications may differ.
  • Basic Weighing Function: Simple on/off, tare, and print functions.
  • Calibration and Adjustment: Internal calibration/adjustment for ATL...-I and ATL...-V models, and external calibration for other models.
  • Configuration (Operating Menu): Allows users to configure the balance to individual requirements, including menu navigation, parameter settings (e.g., balance parameters, interface, printout, extras, application programs, input, info, language), and reset menu.
  • Input: ID number, Date and Time: Allows for setting and displaying ID numbers, dates, and times for ISO/GLP compliant data records.
  • Device Information: Displays software version, serial number, and model designation.

Maintenance Features

Proper care and maintenance are crucial for the longevity and accuracy of the ATILON balances:

  • Repairs: Must be performed by trained service technicians.
  • Cleaning:
    • Unplug the AC adapter from the wall outlet before cleaning.
    • Do not use aggressive cleaning agents (solvents or similar agents).
    • Wipe down the balance with a soft, dry cloth.
    • For analytical balances, remove and clean the weighing pan, drip/breeze ring, and lift it carefully.
    • Clean all stainless steel parts regularly. Use a damp cloth or sponge to clean stainless steel parts. Household cleaning agents suitable for use on stainless steel can be used. Rinse the equipment thoroughly, making sure to remove all residues.
    • To apply oil to cleaned surfaces for additional protection, allow the equipment to dry.
  • Safety Inspection: If there is any indication that safe operation of the balance is no longer warranted, turn off the power and disconnect the equipment from AC power immediately.
  • Recycling: Information on disposal and repairs is provided. The packaging is made of environmentally friendly materials. The equipment, including accessories and batteries, must be disposed of at the local waste disposal facility.
  • Troubleshooting Guide: Provides solutions for common error codes and issues displayed on the main display.

The ATILON series is designed to comply with the following EC Directives and European Standards: Council Directive 89/336/EEC "Electromagnetic compatibility (EMC)" and 73/23/EEC "Electrical equipment designed for use within certain voltage limits."