The A/D converter can now be treated like a standard weighing platform in 
connection with the load sensor. 
t   Close the menu access switch (right position) and reattach the cap.
 
h
  Once ADC configuration has been completed, an adjustment of the weighing 
platform (calibration/adjustment and linearization) must be carried out (see page 47 
and “Calibration/Adjustment without Weights", page 30).
   Entering Geographical Data for Use in Legal Metrology
  Purpose  Entering geographical data allows the external adjustment of weighing equipment 
at a place (e.g. at the manufacturer or vendor's place of business) that is not the 
same as the place of installation. If the weighing equipment is adjusted at the place 
of installation, it is not necessary to enter geographical data. 
The sensitivity of weighing equipment changes depending on the place of 
installation as it is dependent on the on-site gravitational force – or, more precisely, 
on gravitational acceleration. Saving geographical data makes it possible to change 
the place of installation of the weighing equipment after external adjustment has 
been carried out.
The adjustment of weighing equipment is valid at the place of installation and 
within a specific tolerance zone. At 3000 e this zone extends ±100 km from the set 
geographical latitude and ±200 m from the set elevation above sea level.
  Installation Location in Germany  An exception to this is the setting for “Germany (Zone D):” If during external 
adjustment of weighing equipment within Germany the geographical data
–  Geographical latitude: 51.00 degrees N
–  513 m elevation above sea level
are entered, the weighing equipment can be used throughout Germany. 
Gravitational acceleration for “Germany (Zone D)” is 9.810 m/s
2
. On delivery the 
geographical data for “Germany (Zone D)” are entered in the output device.
It is recommended to use the geographical data settings for “Germany (Zone D)” 
when adjusting and delivering the weighing equipment within Germany. Entering 
exact geographical data will lead to a higher level of accuracy but will also restrict 
the tolerance zone.
  Setup information  –   It is only possible to enter geographical data when the menu access switch is 
open. 
–   When the Service mode is activated, you can enter geographical data in the 
Setup menu under “WP-1" for the first weighing platform and COm1 / WP-2, 
UNICOM / WP-2 or COm-WP for the second weighing platform. The settings are 
made in the corresponding Setup menu under menu item 1.20.
–   You can enter either the “geographical latitude in degrees" (LATITUD menu 
item 1.20.1) and “elevation in m above sea level" (ALTITUD menu item 1.20.2), 
or the value for gravitational acceleration (GRAVITY menu item 1.20.3).
   Gravitational acceleration takes precedence over the geographical latitude and 
elevation of the location: If it has been entered, input fields for latitude and 
elevation show the values 99999.99 and 9999999 respectively. If only elevation 
and latitude have been entered, 0000000 is displayed for gravitational 
acceleration.
 
3
   If you return to the highest level of the Setup menu without saving the configura-
tion parameter beforehand (save menu item 1.20.4) any settings that have been 
made will be deleted. 
24  Operating Instructions Combics Indicators
  Configuring Weighing Platforms