Setting the Preload
Setup information Setting the preload when weighing in legal metrology is only possible when
the menu access switch is open.
The “Set Preload" function (menu item 1.9.8) must be allocated to the J
key (see page 141).
3
Once the preload has been set, the J key must be reallocated back to its
original function in the Setup menu, e.g. external calibration/adjustment with
default weights (Setup menu item 1.9).
Procedure
( t Zero the weighing platform.
t Place the preload weight on the weighing platform.
J t Start the “Set Preload" function.
y After a short period of time the preload will be applied and the indicator
will automatically switch back to weighing mode.