Setting: Applic./Applic.1/PERCENT/Min.init menu item 3.6.
The minimum load can be set in 10 steps from 1 to 1000 digits (see available 
parameters). The “digits" here refer to the scale intervals for the connected weighing 
platform. If the interval of the connected platform is 1 g, for example, and 1000 
digits are required, you must place at least 1000 g (=  1000 intervals =1000 digits) 
on the weighing platform for initialization.
 Resolution The resolution indicates the accuracy used to determine the reference weight. 
The default setting is “display resolution." The resolution is increased when “10-
fold" or “100-fold" is selected. “10-fold" increases the resolution of the net value 
by one step (display resolution x 10), “100-fold" increases it two steps (display 
resolution x 100).  
Setting: Applic./Applic.1/PERCENT/RESOLUT menu item 3.9.
  Decimal Places  The number of decimal places displayed can range from none up to 3 places.  
Setting: Applic./Applic.1/PERCENT/DEC.PLCS menu item 3.10.
  Parameter for saving weight values  The weight on the platform is saved as a reference value as soon as the platform has 
stabilized. “Stability" is defined as the point at which the fluctuation of a measured 
value lies within a defined tolerance. The narrower the tolerance, the more stable the 
platform is at “stability".
The “increased stability" setting has a lower tolerance so that the average piece 
weight saves is more accurate and the results more reproducible; however, the 
measurement time can take longer.  
Setting: Applic./Applic.1/PERCENT/SAVE WT. menu item 3.11.
  Weighing in percent with   You can use two weighing platforms simultaneously with the Weighing in Percent
  two weighing platforms   application. When using two platforms, you can choose from the following 
operating modes:
–  Weighing in percent with two platforms of the same type
–  Weighing in percent with one reference platform and one weighing platform
Weighing in percent with two platforms of the same type
Use this mode to measure different types of sample material with different weights. 
For example, measure the lighter-weight samples on one platform and the heavier 
samples on another. You can define one of the two scales as the reference scale. 
The reference scale is the first scale active when you switch on the device, regardless 
of the setting for automatic initialization of the Neutral Measurement application.
Weighing in percent with one reference platform and one weighing platform
In this operating mode, the reference platform is a high-resolution weighing 
platform with a relatively low maximum capacity. The other platform is used for 
weighing heavier samples, and has a high capacity with a relatively low resolution.
This allows you to both determine the reference piece weight with high resolution; 
i.e., very precisely, and to measure large samples, without requiring an expensive 
high-resolution, high-capacity weighing platform.
The system can be configured to switch automatically to the reference platform for 
initialization (the measured value line shows Ref). Following initialization, you can 
switch to the counting platform.
  Operating Instructions Combics Indicators  69
  Operation