) t Tare the scale.
This is not required if the automatic tare function is active. The tare weight is
saved automatically when you place the container on the platform.
y The prompt to fill and save the first component is shown.
t Place the first component into the container (in this example, 1100 g).
y The weight of the first component is displayed.
O t Save the weight of the first component using the O key.
Cmp001+ 1.100 kg yThe component record is printed automatically.
y The weighing platform is tared and the component counter value is increased by
one. The prompt to fill and save the second component is now displayed.
t Place the second component into the container (in this example, 525 g).
y The weight of the second component is displayed.
O t Save the weight of the second component using the O key.
Cmp002+ 0.525 kg yThe component record is printed automatically.
y The weighing platform is tared and the component counter value is increased by
one. The prompt to fill and save the third component is now displayed.
w t Toggle to the “additive mode” using the w key to display the total weight of
all components.