GMP-compliant Printouts
When the corresponding menu item is
active, the measured result is bracketed
on the printout by a GMP header and
a GMP footer (GMP = “Good Manufac-
turing Practice").
The GMP header precedes the first
measured result. The GMP footer is
printed either after each measured
result (“ISO/GLP/GMP: For 1 application
result," menu item 7.11.2), or after the
last result in a series of measurements
("ISO/GMP/GLP: For several application
results," menu item 7.11.3). To end a
series of measured results, press and
hold the p key (> 2 sec). In this case,
the A symbol is displayed after the
GMP header is printed and remains in
the display until the GMP footer is
printed.
If you toggle to a different weighing
instrument (Combics 2 only) while
a GMP printout of several measured
results is being generated, the GMP
footer for the weighing instrument used
up to that point is generated when
you press n. The GMP header for the
other weighing instrument is included
on the next printout generated.
A GMP-compliant printout is generated
automatically at the conclusion of
calibration/adjustment routines, as well
as when you set or clear a preload.
If you use a label printer for GMP-
compliant printouts and menu item
7.11.3 is active, the header and footer
are printed on two different labels.
To generate GMP-compliant printouts
on labels, select menu item 7.11.2.
Three examples of GMP headers and
one example of a footer are shown in the
following. On Combics 1 models, the
“date and time" line is not included.
Weighing platform WP1:
-------------------- Dotted line
14.01.2002 09:43 Date/time
1
)
Model CISL2 Combics model
Ser.no. 12345678 Combics serial no.
Vers. 1.1007.12.1 Software release for applications
BVers. 01-25-01 Software release for basic version
-------------------- Dotted line
Weighing platform WP2 (xBPI protocol):
2
)
-------------------- Dotted line
14.01.2002 09:45 Date/time
1
)
Model CISL2 Combics model,
Ser.no. 12345678 Combics serial no.
Vers. 1.1007.12.1 Software release for applications
BVers. 01-25-01 Software release for basic version
Type IS12000S Platform model
Ser.No. 12345678 Platform serial no.
-------------------- Dotted line
Weighing platform WP2 (SBI protocol):
2
)
-------------------- Dotted line
14.01.2002 09:45 Date/time
1
)
Model CISL2 Combics model
Ser.no. 12345678 Combics serial no.
Vers. 1.1007.12.1 Software release for applications
BVers. 01-25-01 Software release for basic version
Type SBI (Platform model)
-------------------- Dotted line
GMP footer:
-------------------- Dotted line
14.01.2002 09:45 Date/time
1
)
Name: Field for operator signature
Blank line
-------------------- Dotted line
1
) Not applicable for Combics 1 indicators
2
) Combics 2 indicators only
100
Data Interfaces